Perchloric Acid Fume Hood: Operation and Safety Guide

2026-09-13

Is Perchloric Acid a Strong Acid?

Perchloric acid is the strongest inorganic acid and also the most acidic acid in the world at present and  it is highly corrosive. Perchloric acid explosion might happen when the temperature exceeds 130℃. Perchloric acid becomes much more stable after being absorbed by perchloric acid fume hood with washdown system.

Perchloric acid can accumulate in the exhaust system to form sodium perchlorate. It may explode when exposed to uncertain factors such as high temperature, vibration, and organic matter. Therefore, the spray system of perchloric acid fume cupboard is applied to perform a cleaning function of no less than 5 meters on the exhaust duct. Sodium hydroxide is usually used as the absorption and neutralization liquid, with a solution concentration of 2-6%, and the purification efficiency is all above 90%.

So the operation and safety guide is very importent during the whole running processes of perchloric acid fume hood. This article describes how to operate the fume hood in different operation phases, precautions, first aid measures and maintenance. It would give you some practical and useful tips and help you use the equipment safely when you operate it.

safety and operation guide perchloric acid fume cupboard

Occupational Health Protection and Treatment

Operation precautions: closed operation, local ventilation.

The operation shall be mechanized and automatic as far as possible. Exercising operators must be specially trained and strictly abide by the operating procedures. It is recommended that operators wear filter type anti lock gas mask (full face mask) or self-contained breathing apparatus, wear polyethylene gas suit and rubber gloves.

Keep away from fire and no smoking is allowed in the workplace. Prevent steam from leaking into the air of the workplace. Avoid acid Class A, alkali and amine contact. In particular, avoid contact with water. Handle with care to prevent damage to packaging and containers. Vibration, impact and friction are prohibited. Fire fighting equipment and leakage emergency treatment equipment of corresponding varieties and quantities shall be provided. Empty containers may retain harmful substances. When diluting or preparing solution, add acid into water to avoid boiling and splashing.

Circulating Water Dilution System

purification system of perchloric acid fume hood

There may be an aerosol release process of perchloric acid in the experiment conducted in the perchloric acid fume hood

The fume hood needs to exhaust under the condition of rated large air volume, and the design value is 1300CMH for 1500mm wide percholoric acid fume cupboard.

The perchloric acid is mixed with the alkaline water behind the draft plate of the fume hood, and the chemical reaction is:

HClO4+NaOH====NaClO4+H2O

The neutralized product is neutral, and its PH value is 7-14, so the circulating treatment in this design.

The system is made of chemical resistant materials.

A buffer tank is set at the lower part of the fume hood, and PP water pump is connected at the outlet side. The water pump and the fog behind the deflector

The chemical pipe is connected, and the water pump outputs the alkali liquor to the atomization pipe in the form of medium pressure, which forms fine water after passing through the atomization pipe

The droplet is mixed with the gas of perchloric acid to complete the chemical reaction. Under the influence of gravity, the product passes through

Re enter the buffer tank of the water pump through the technical interlayer for circulation.

Pre-Checking before Operation Processing System

It including the following:

1) The water tank shall be filled with alkaline water (solution with 50% concentration of NAOH), and the current PH shall be monitored by the PH detector

2) Check whether all valves in the circulating water system are open. The circulating pump shall not be started if any valve is closed

3) Whether perchloric acid test is prepared in the fume hood, and the fume hood shall have rated exhaust air volume (The face velocity of the area is 0.5-0.8m/s, and the hand-held anemometer is required for detection)

4) Turn on the start button of the water pump, and the circulating water is evenly distributed behind the deflector in the form of water droplets.

Safety Operation Process after Experiment

1) The perchloric acid test in the fume hood has been completed

2) Close the water pump, and record the PH value through the PH detector for reference value preparation for the next experiment.

(When the PH value is below 6, the buffer solution needs to be replaced)

3) Close the valve on the water system

4) Close the ventilation of the fume hood

controling system of perchloric acid fume hood

Precautions of Perchloric Acid Fume Cupboard

PH instrument setting: PH high alarm is set as 11-13, and low alarm is set as 4-6.

This type of fume hood is a customized perchloric acid fume cabinet specially designed for perchloric acid. The user is requested to use it according to the characteristics of perchloric acid

Carry out the experiment according to the characteristics. During the experiment, it is necessary to follow the relevant experimental operation procedures. Do not directly conduct the experiment in the fume hood

High concentration and large dose perchloric acid experiments were conducted. The alkali water neutralization circulating system behind the draft plate of the fume hood is a primary neutralization system, which is used to ensure that perchloric acid residues with high concentration will not be accumulated in the fume hood, thus causing combustion, explosion or a large number of heating chemical reactions. It does not guarantee that the relevant national exhaust emission standards can be met. In order to meet the national environmental protection emission standards, secondary neutralization reaction is required at the fan side.

First Aid Measures

Skin contact: Take off contaminated clothes immediately, and wash with plenty of flowing water for at least 15 minutes. Get medical attention.

Eye contact: immediately lift the eyelids and wash them thoroughly with a large amount of flowing water or normal saline for at least 15 minutes Clock. Get medical attention.

Inhalation: quickly leave the site to a place with fresh air. Keep the respiratory tract unobstructed. If breathing is difficult, give oxygen. as stop breathing, and perform artificial respiration immediately. Get medical attention.

Ingestion: Rinse with water and drink milk or egg white. Get medical attention.

If you are interested in knowing working principle, site requirements, technical sepcifications and main components of fume hoods for perchloric acid, treatment principle about installation, pls visit  technical overveiw of perchloric acid fume hood.

Maintenance of Concentrated Perchloric Acid Fume Hood

Structural schematic diagram of circulating pump

Structural schematic diagram of circulating pump

In case of leakage of the water pump, please stop the test immediately, and contact the contractor for timely maintenance;

The circulating water treatment system needs to regularly clean the sludge precipitation in the buffer tank;

The water pump shall not idle without liquid;

The water pump shall not operate when the pipeline valve is not opened;

The water pump shall not operate with the chemical solution whose temperature exceeds 50 ℃;

 

Maybe you like also

  • Categories

  • CONTACT US


    Questions?

    0086-0510-83888695

    info@laboratoryfurnitures.net

    0086-15370203085

    noicle1

     

  • OUR COMPANY

    JIAHONG LABORATORY is a professional lab furniture and fume hood manufacturer & Supplier from China. We specialize in laboratory furniture and turnkey lab solutions.

  • CONTACT US

    0086-0510-83888695

    info@laboratoryfurnitures.net

    0086-15370203085

    noicle1

    Exclusive Representative In Brasil